FE-707-EC Error Codes
Error codes
Configuration errors
These occur at startup only.
| Value | Function | Notes |
|---|---|---|
| 20 | Flash block checksum failure | New |
| 21 | RAM block checksum failure | New |
| 22 | Flash & RAM block checksum failures | New |
| 53 | Module unavailable or changed |
General errors
| Value | Function | Notes |
|---|---|---|
| 30 | Comms syntax error | New |
| 32 | Invalid channel number | |
| 33 | Invalid group number | |
| 34 | Invalid parameter value | Includes legacy error 31, not used |
| 69 | Invalid host command |
Module-specific errors
Value errors occur if a write parameter is out of range.
Write or read errors occur if the feature is not fitted on the selected module.
| Error type | Parameter | ||
|---|---|---|---|
| Value | Write | Read | |
| 35 | 46 | 58 | Filter Cutoff Frequency |
| 36 | 47 | 59 | Low-Pass Filter Frequency |
| 37 | 48 | 60 | High-Pass Filter Frequency |
| 38 | 49 | 61 | Bridge Supply |
| 39 | 50 | 62 | Gain Level |
| 40 | 51 | 63 | Calibration Shunt Resistance |
| - | 52 | 64 | Auto Zero |
| 42 | 54 | 65 | Calibration Voltage Injection |
| 43 | 55 | 66 | External Calibration |
| 44 | 56 | 67 | AC/DC Coupling |
| 45 | 57 | 68 | Constant Current/Voltage |